Specifications

Available Standard Versions
Performance Specifications
Isolation technology:
Control electronics:
Force directions:
Isolation performance:
Active bandwidth:
Settling time:
Max. correction forces:
Load capacity:
Other Specifications
Dimensions:
Weight:
Maximum compensation
level:
Interface:
Environmental and Operational Requirements
Electrical voltage:
Power consumption:
Operating temperature:
Relative humidity:
Operating altitude:
Certification
Electrical Safety:
EMC:
*
Consists of four isolation elements and a 4 port control unit.
** Floating table top is supported by steel springs; low-pass characteristics of spring-mass
combination dominates the dynamic behaviour above 200 Hz.
*** The maximum compensation level depends on several conditions, such as payload, frequency, load
distribution and height of the payload as well as the distance between the isolation elements.
For that reason this value should be considered as an estimation.
Vario Basic 40-100
Vario Basic 60-100
Vario Basic 90-100
Halcyonics active vibration isolation technology based on piezoelectric type
acceleration pickup, fast signal processing and electro-dynamic force trans-
ducers.
External control unit with sensor and actuator LEDs, corresponding to force
directions
Active compensation in all six degrees of freedom
> 5 Hz = 25 dB
(94.4%)>10 Hz = 38
dB (98.75%)
0.7 – 200 Hz**
100 ms
Vertical ± 8 N;
horizontal ± 4 N
0 – 100 kg (0 - 220 lbs) 0 – 300 kg (0 - 660 lbs) 0 – 600 kg (0 – 1,320 lbs)
See page 11
Vario Basic 40: 6,8 kg (15 lbs) per isolation element
Vario Basic 60: 8,6 kg (19 lbs) per isolation element
Vario Basic 90: 10,4 kg (23 lbs) per isolation element
Vario Basic control unit: 4,5 kg (10 lbs)
550 µm/s at 6 Hz and
60 kg (132 lbs)***
BNC analog diagnostic output – 50 Ohms
115 - 230 V AC / 50 – 60 Hz
Typically 10-20 W; max. 60 W
10 – 40°C (50 – 104 F)
0 – 60%
< 2500 m (8100 ft)
CE certificated according to directive 2014/35/EU
CE certificated according to directive 2014/30/EU
Vario Basic 40-300
Vario Basic 60-300
Vario Basic 90-300
> 5 Hz = 25 dB
(94.4%)>10 Hz = 35
dB (98.2%)
1.0 – 200 Hz**
300 ms
Vertical ± 8 N;
horizontal ± 4 N
500 µm/s at 8 Hz and
150 kg (330 lbs)***
Vario Basic 40-600*
Vario Basic 60-600*
Vario Basic 90-600*
> 5 Hz = 25 dB (94.4%)
>10 Hz = 35 dB (98.2%)
1.0 – 200 Hz**
300 ms
Vertical ± 16 N;
horizontal ± 8 N
500 µm/s at 8 Hz and
300 kg (660 lbs)***
- 11 -