Accuris ASPIRE Manuel de l'utilisateur - Page 6

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Accuris ASPIRE. Accuris ASPIRE 8 pages. Laboratory aspirator

The collection bottle and lid can be autoclaved by placing the parts into an autoclave and running one sterilization
stem cycle at 121°C (1 bar relative pressure) making sure that the jar is positioned upside down.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated
conditions (EN ISO 10079-1). Beyond this limit the physical-mechanical characteristics of the plastic may decrease
and replacement of the part is therefore recommended.
After sterilization and cooling at ambient temperature, make sure that the parts are not damaged.
Assemble the bottle as follows:
Place the overflow valve into its seat in the cover (under VACUUM connector)
Insert floating valve keeping the o-ring towards the opening of the cage
Place the o-ring into its seat around the cover
After completing assembling operations always make sue that cover seals correctly to avoid vacuum or
liquid leakage.
The aspiration tubes can be sterilized in an autoclave using a sterilization cycle at 120°C.
DO NOT WASH, STERILIZE OR AUTOCLAVE THE ANTIBACTERIAL FILTER
The ACCURIS ASPIRE LABORATORY ASPIRATOR does not require any routine maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. Unpack the instrument and always check the plastic
parts for any damage that may have occurred during prior use. Connect cable to an appropriate electrical outlet
with correct voltage and turn the power switch to on.
Close the aspiration outlet by blocking with a finger and with suction regulator set to maximum check that the
vacuum gauge reaches at least -75 kPa (-0.75 bar) maximum. Rotate the knob from right to left and check the
aspiration regulating control.
The vacuum indicator should go down to -25 kPa (-0.25 bar). Check that no loud noises are present.
A fuse (F 1x4A 250V) is located in the plug to protect the instrument. In the event of a fuse blowing, replace with
another or the correct type.
Fault type
1.
The vacuum system
doesn't work
2.
No aspiration
3. No aspiration
4.
The Vacuum power
on the sample side is very
low or absent

MAINTENANCE

Cause
Cable is damaged
External power source failure
Bottle Cap not screwed down
properly
Silicone cap seal (o-ring) not
seated properly
a) Vacuum regulator set to
minimum
b) Protection filter blocked or
damaged
c) Connection tubes blocked,
kinked or disconnected
d) Shut-off valve blocked or
damaged
e) Pump motor damaged
Replace the cable
Check the external power source
Unscrewed the cap, and re-attach carefully
Unscrew the cap and insert the seal properly.
a) Turn the vacuum regulator clockwise and check the
value of the vacuum on the gauge
b) Replace the filter
c) Replace or reconnect the tubes, check the bottle
connections
d) Empty the bottle, or disconnect the tube from the
bottle and unblock the shut-off valve. The unit twill
only work in the upright position.
e) Refer to authorised service personnel
6
Solution