5. Wait 5 to 8 seconds, and 200.0000 value will be displayed. The 200g
calibration is complete.
Check the linearity at different weight values to confirm proper
calibration. If the results are outside of specification, repeat above
steps.
5. FUNCTION SETTINGS

5-1 CHANGING FUNCTION SETTINGS

You can reset and change the function as follows.
 Turn on the balance with the weigh pan empty;
 Wait for the display to show 0.0000g
 Press the SET key then the PRINT key to enter the FUNCTION
SETTING MODE.
 C1- - - 0 will be displayed.
 Press the TARE key to scroll through the different available
FUNCTIONS from C1 through C9.
 Press the PRINT key to change the FUNCTION setting numbers.
 When finished changing a FUNCTION SETTING, press the TARE
key, and the display will show the next FUNCTION.
 Press the Power Button, and "SAVE _ _ _ " will be displayed.
 Press the TARE button, and "S - - End" will be displayed.
 Press On/Off.
The new function settings are now stored in memory.

5-2 FUNCTION SETTING INDEX

FUNCTION
C1:Calibration
mode
C2:Set the
reference number
of samples for
piece counting
SETTING
C1—00
Auto-calibration
weight
C1—01
Manual calibration using external weight
C2—0
10
C2—1
20
C2—2
50
C2—3
100
C2—4
1000
C3—00
No "0" point tracking status
12
DETAILS
using
the
internal