Siemens SITRANS F M Verificator Manuel d'utilisation - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Siemens SITRANS F M Verificator. Siemens SITRANS F M Verificator 38 pages.

Siemens SITRANS F M Verificator Manuel d'utilisation
0%- ##
Press ESC to cancel
Transmitter test finished
Press ESC to cancel test.
T1:3 Transmitter test
0%-##############
Press ESC to cancel
Is only displayed if the transmitter failed the test.
Press ESC to quit as a complete test requires a transmitter which have passed.
See On site troubleshooting (Page 33)
Replace the transmitter and recommence the test.
E1: Transmitter failed
Press ESC to cancel
Is only displayed if the Verificator failed the digital output on the transmitter.
Replace the transmitter and recommence the test.
T1: Transmitter test
0%-##############
Digital Output Error
Press ESC to cancel
Transmitter passed the test.
The verification will continue automatically.
I1: Transmitter passed
Please wait .......
Is only displayed if the SENSORPROM
Press GO to finish the verification test.
The SENSORPROM
Verification finished
Transmitter:
SITRANS F M Verificator
Operating Instructions, 07/2019, A5E45134218-AB
-100%
-100%
-100%
unit is not installed, or defect.
®
unit can be checked with the SENSORPROM
®
-100%
Operating (hardware)
3.2 Verificator software - transmitter
programmer .
®
11