Operating (hardware)
3.6 Verificator software - part test
Is displayed when the test is passed.
I3: Sensor Passed
Magnetism OK
Press GO to continue
If the test fails you can get the following messages
Is displayed if the magnetic circuit is open.
Verification finish
Sensor:
Open circuit
Press GO to continue
Is displayed if the measured values are out of range.
Verification finish
Sensor:
Out of range
Press GO to continue
Is displayed if the gathered readings are unstable.
Verification finish
Sensor: Selftest Fail
Unstable reading
Press GO to continue
16
Selftest Fail
Selftest Fail
SITRANS F M Verificator