600Ω/
/
Set the slide-switch function selector to the 600Ω/
●600Ω Resistance range with Audible-Continuity is the default function. It is an
extended low resistance range to complement the Resistance (Ω) function in
AutoCheck
mode. Audible-Continuity response time is also improved drastically
TM
(from that of AutoCheck
Audible-Continuity is convenient for checking wiring connections and operation of
switches. A continuous beep tone indicates a complete circuit.
●Press SELECT button momentarily selects Diode test function. The reading shows
the approximate voltage drop across the test leads. When forward biased, normal
forward voltage drop for a good silicon diode is between 0.400V to 0.900V. A reading
higher than that indicates a leaky diode (defective). A zero reading indicates a shorted
diode (defective), and the meter gives a continuous beep warning. An OL indicates an
open diode (defective). Reverse the test leads connections (reverse biased) across the
diode. The display shows OL if the diode is good. Any other readings indicate the diode