My readings are erratic
My SNAP is really noisy
My Snap can't seem to run wild swinging
all over the place, and can't seem to hit a
set point.
My Snap seems to be "breathing"
©
2021 DigiVac Company®Allrights reserved. | 1020 Campus Drive, Morganville,NJ 07751 | (732) 765-0900
Check reading with another gauge to see
if the readings are indeed erratic. If the
other gauge does not show erratic
readings, consult your vendor.
The valves might make a whiny or
buzzing sound at startup. This is normal.
Adjust PID variables, probably reduce
variable "P"
The base pressure of the pump is probably
high as well. This can be solved by resetting
the PID. If SNAP is controlling too
aggressively, try setting P=p/2. Want to learn
more see our white paper on PID