Press [FUN] to confirm and go to label format selection. Options: 1/2
"P11 FL": Weight adjustment for acceleration of gravity change.
"P12 KB": on: keyboard beep is on; off: keyboard beep is off
"P13 UT": Unit Setting.
Default unit: lb/kg
Lb on/off
U1 on/off
U2 on/off (Unit 1 or Unit 2 setting)
"P14 Pn": Password Setting.
P1: Enter new password
P2: Re-enter the new password
Enter Parameter setting P8.
"Unload": Empty the scale platform. Wait until the STAB sign is on, press [FUN].
Input the calibration weight value and press [FUN]
"Load": Place the calibration weight on the platform, wait until the STAB sign is on, press [FUN]
The indicator will go to weighing mode when the calibration is completed.
Chapter 6 Error Indication
--OL--
Err 1 / Err 2
Err 4
Floor / Pallet Scale
Press [FUN] to confirm.
Press [FUN] to confirm. Screen will show "done".
Chapter 5 Calibration
The scale is overloaded. Empty the scale platform.
Date and Time are not right. Reset the date and time.
The zero point is out of range upon powering-on.