IT-8 Back cover:
Key-press definition:
"SET" key: using for parameter setting, on the state of waiting or working, press the key
for 5 sec, enter the state of parameter setting , on the course of parameter setting, press the
key for 5 sec, back to the state of temperature displaying. Without any action for 15 min, the
display will shut down automatically. On the state of blank screen, pressing the key can
activate the temperature display.
" "key:On the state of parameter, using the key for moving radix point, meanwhile, the
position of radix point effect.
" "key: Using for program the position of selected.
Parameter setting manual table
Parameter Meaning
E1
Emissivity 1
E2
Emissivity 2
AHd
Peak
keeping time
dEFd
Difference
keeping time
CoEF
Filter coefficient
Ad
Interface address
rAd
Measuring mode
4 Model naming and selection
(1)Model naming
IT– X X X X
Wuhan Dikai optoelectronics technology Co., ltd
Address: No 709, street of Xiongchu, Wuhan
Tel: +86-27-87509386 87509105
WUHAN DIKAI OPTOETECTRONIC TECHNOLOGY. Co., LTD
Range
0.01~1.00
0.01~1.00(ITW only)
value
0~120 sec
0~120 sec
value
0~20
1~100
0:Continious measuring;1:Peak
measuring ;2:Difference measuring
English Letter output sorts V:0~5V;I:4~20mA
Number Temperature range 1: 400℃~1200℃
Number Accuracy level
Number Series code
English abbreviate(Infrared-Thermometer)
1:0.5 level
6: no laser collimation;
8: laser collimation
Post code: 430073
Fax: +86-27-87509386
http://www.dkoe.com
Default
1.00
1.00
0
0
0
1
0