MECHANICAL SPECIFICATIONS
Operating Temperature:
(-5 to 122) °F (-15 to 50) °C
Storage Temperature:
(-40 to 185) °F (-40 to 85) °C
Influence of Temperature:
< 0.1 % per 10 °K
Altitude:
Operating: (0 to 2000) m
Non-operating: (0 to 12,000) m
Jaw Opening:
1.3 in (33 mm)
Maximum Conductor Size:
1.18 in (30 mm)
Maximum Bus Bar Size:
(2.48 x 0.20) in (63 x 5) mm
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
1.5 m (IEC 68-2-32)
Mechanical Shock:
100 g (IEC 68-2-27)
Vibration:
10/55/10 Hz, 0.15 mm
(IEC 68-2-6)
Polycarbonate Material:
Handles: 10 % fiberglass charged
polycarbonate UL 94 V0
Dimensions:
(2.6 x 7.68 x 1.34) in (66 x 195 x 34) mm
OPERATION
Please make sure that you have already read and fully understand the WARNING section on
page 1.
Making Measurements with the AC Current Probe Model MD304
•
Connect the black and red terminals to the Ampere AC range of your DMM or current
measuring instrument. Select the appropriate current range (10 A
probe around the conductor to be tested with the arrow pointed toward the load. Read the
value display on the DMM and multiply it by the probe ratio (100/1). (If reading = 4.59 A,
the current flowing through the probe is 4.59 A x 100 = 459 A
•
For best accuracy, avoid if possible, the proximity of other conductors which may create
noise.
Tips For Making Precise Measurements
•
When using a current probe with a meter, it is important to select the range that provides
the best resolution. Failure to do this may result in measurement errors.
•
Make sure that probe jaw mating surfaces are free of dust and contamination.
Contaminants cause air gaps between the jaws, increasing the phase shift between
primary and secondary. It is very critical for power measurement.
Weight:
14.82 oz (420 g)
Colors:
Dark gray handles
Output:
Insulated 5 ft (1.5 m) lead with safety 4 mm
banana plugs
SAFETY SPECIFICATIONS
Electrical:
Double insulation or reinforced insulation be-
tween the primary or secondary and the outer
case of the handle per IEC 1010-2-032.
- 600 V CAT III, Pollution: 2
- 300 V CAT IV, Pollution: 2
Electromagnetic Compatibility:
EN 50081-1 Class B
EN 50081-2 Electrostatic discharge
IEC 1000-4-2
Radiated field IEC 1000-4-3
Fast transients IEC 1000-4-4
Magnetic field at 50/60 Hz IEC 1000-4-8
ORDERING INFORMATION
Current Probe MD304 ..............Cat #1201.37
Accessories:
Banana plug adapter
(to nonrecessed plug) ................. Cat #1017.45
AC
range). Clamp the
AC
)