6.2.
Service and Repair
Direct connection flowmeters do not require routine servicing but can be repaired. The device can
be returned to a recognised BPR Medical Service Centre, details of which can be found at
www.bprmedical.com.
Indirect connection flowmeters should have the hose assembly replaced after four years use or
five years from the date of manufacture. The manufacture date can be found on the device label.
Device life can be calculated by adding 7 years to the date of manufacture.
Disassembly, repair or reassembly must be carried out by a suitably qualified person working in a
controlled environment.
7.
Specification
Specification
Flow Scale
Inlet Pressure Range
Supply Pressure Range
Accuracy
Gas Compatibility
Maximum Flow (Flush Flow
Mode)
Firesafe™ Nozzle Flow Range
Inlet Filtration
Environmental
Regulatory
BS EN ISO 5359
BS EN ISO 15001
BS EN ISO 15002
BS EN ISO 19054
0 – 15 l/min
Model specific between 370 – 550 kPa
Note: The flowmeter is able to operate safely within the pressure range 370 kPa
to 550 kPa, but the flowmeter will only deliver accurately when connected to an
input pressure as shown on the device. Any input pressure other than shown on
the device will result in flow rates different from the flow settings
Minimum 370 kPa
Maximum 550 kPa
Nominal Pressure: ±10 % of indicated reading between 1 – 15 l/min
Valve Settling: < 2 % from initial setting within 3 hours
Temperature: 3 % flow increase/decrease per 5 °C temperature increase/decrease
Inlet Pressure: 4 % increase/decrease per 10 kPa increase/decrease from nominal
pressure
Outlet Flow Resistance: Negligible up to 230 kPa gauge back pressure
Medical Oxygen or Medical Air (model dependant)
25 – 30 l/min
1 – 25 l/min
25 µm
Transport and Storage Temperature: -20 °C to 60 °C (-4 °F to 140 °F)
Operating Temperature: 0 °C to 40 °C (32 °F to 104 °F)
Humidity: 0-100% RH non-condensing
CE: Medical Device Directive 93/42/EEC – Active Medical Device – Class IIa
Applied Standards
Low pressure hose assemblies for use with medical gases
Anaesthetic and respiratory equipment. Compatibility with oxygen
Flow-metering devices for connection to terminal units of medical gas pipeline
systems
Rail systems for supporting medical equipment
Value
8