• Turn on the computer and start the EC calibration station.
• Calibration station toolbox opens.
• Go into top right-hand corner icon with hand and paper.
• The system settings page is displayed.
• Go into Test Setup Page
• Go to drop down bar on Theme to select recalibration mode 'Recal Sleep'
• Close the System Settings window
• Choose 'Connect' Icon at top left of EC Calibration Station window System will
automatically register a unit that has been in the factory historically
o Check serial number is the same on both sticker and EC Calibration window
o If the system does not recognize the instrument, a message will appear in a
separate window indicating "New Unit"
o When the system indicates "New Unit" the Edit Unit Data screen will appear
automatically.
• This information can only be updated once the testing has begun.
REF: EC Service and Calibration Instructions V8
Click on this icon to display System
Setting setup screens required for
following operation step procedure.
Page 5