12.2 Replacing sample module

Rather than returning the instrument to Alcolizer for service, the module can be removed
and sent in. This saves money and if the owner details are known and maintained with
Alcolizer Technology, arrangements can be made to forward a replacement sample module
prior to the calibration being overdue.
Do not over tighten the sample module securing screw as this will damage the threaded
portion. Tighten only enough to prevent the sample module from sliding away from the
Remove and replace the sample module as follows:
a. Undo the single screw on the back of the main module.
b. Slide off the out-of-calibration sample module.
c. Slide on the new sample module.
REF: Druglizer LE5 Drug Tester User Manual V7.docx