1 INTRODUCTION

The Druglizer LE5 Drug Testing System is for the rapid oral fluid screening of drugs of abuse.
The system incorporates a Druglizer LE5 Drug Tester Cartridge (Druglizer LE5 Cartridge), a
Druglizer Oral Fluid Collector and a Druglizer LE5 Drug Tester instrument (Druglizer LE5,
purchased separately). The system is intended to be used by a trained professional to
provide a first line screen. A positive result displaying 'Unconfirmed' on-screen will require
confirmation by a laboratory-based reference assay like Gas chromatography – Mass
Spectrometry (GCMS) or Liquid Chromatography Mass Spectrometry (LCMS).
This document must be read carefully before using the Druglizer LE5 Drug Testing System
and Certified Operator Training provided by Alcolizer Technology must be completed before
using the Druglizer LE5 Drug Testing System. Keep the user manual available for the life of
the instrument and ensure operators have access to the user manual at all times. You must
always hand the user manual over to the subsequent instrument owner or user.
WARNING
Only human oral fluid should be used.
Results on the Druglizer LE5 are preliminary qualitative analytical test results. A
result displaying 'Unconfirmed' on-screen will require confirmation by a laboratory-
based reference assay.
The presumptive presence of drugs is not a measure of impairment.
A negative result may not necessarily indicate a drug free sample. Drugs may be
present below the stated cut off.
Foods/medicines or preclinical errors may result in false results.
The Druglizer LE5 must be stored at a temperature between –10°C and +60°C. Conduct drug
screening tests at a temperature between +15°C and +35°C.
This User Manual is applicable to the Druglizer LE5 with main module firmware version
REF: Druglizer LE5 Drug Tester User Manual V7.docx
Uncontrolled when printed
NOTE
RN-1.5.0 and drug test module or later