Alcolock L OEM Manuel d'étalonnage - Page 3
Parcourez en ligne ou téléchargez le pdf Manuel d'étalonnage pour {nom_de_la_catégorie} Alcolock L OEM. Alcolock L OEM 8 pages. Alcohol interlock, alcohol gas standard
CAUTION!
• It is recommended calibration be done indoors, in a service facility, by maintenance code or
using an ACS download station
• This calibration guide is intended for use by trained technicians only
COMPRESSED GAS SAFETY
• Damaged or broken valves can turn a canister into an unguided missile. Attach the valve in a
safe location
• Calibration should be performed indoors, in a service facility, where the gas standard cylinder
can be properly stored
• Transportation and storage of compressed gases in vehicles is dangerous and should be
avoided
• Examine the canister and valve for any damage, and the expiration date on the label
• Observe all cautions and safety warnings found on the canister
• Never remove or alter canister labels
• Never modify the delivery tube in any way
• Always remove the valve and install the protective cap on cylinders when not in use
• Store cylinders in a cool, well ventilated area away from sources of heat
EQUIPMENT
• 130 PPM (0.05%BAC) alcohol gas standard, ACS #95-000425 or 260 PPM (0.10%BAC)
alcohol gas standard, ACS #95-000436
• Regulator valve assembly, ACS #79-005546
• V3 download station, ACS #79-006111
• V3/L OEM download station adapter, ACS #79-009115
CYLINDER SETUP
Remove the protective cap and screw the regulator valve to the canister. Continue to the next
section.
CALIBRATION MANUAL
1