Alcolock V3 Manuel d'étalonnage - Page 4
Parcourez en ligne ou téléchargez le pdf Manuel d'étalonnage pour {nom_de_la_catégorie} Alcolock V3. Alcolock V3 8 pages. Alcohol gas standard
CALIBRATING A HANDSET
Calibration is possible both with a download station or by maintenance code (without
a download station).
Do A or B.
A. WITH A DOWNLOAD STATION
1.
Connect the download station to a power source.
2.
Disconnect the handset from the ECU.
3.
Connect the download station to the handset with a coiled cable.
The handset automatically powers on.
4.
Press the left button on the handset to select Cal.
The current handset date and time are displayed.
5.
In the Date / Time screen, do I or II:
I.
If the date and time are correct, press OK – the Select method menu is displayed.
II.
If the date or time is incorrect, press Set to automatically synchronize handset time
with download station time, and press OK – the Select method menu is displayed.
NOTE: If a tachograph is installed, the date and time cannot be changed.
6.
In the Select method menu, press Next to scroll to Manual and press Select.
7.
In the Select type menu, press Next to scroll to Dry Gas and press Select. The Select
value menu is displayed.
Skip the next section and read: "Calibrating a handset (continued)."
B. BY MAINTENANCE CODE (NO DOWNLOAD STATION)
NOTE: The handset must be connected to the ECU.
1.
Power the handset on by holding down either the right or left navigation button.
2.
Access the menu screen by pressing and holding the left button.
NOTE: Where available, press and hold the right navigation button to exit a current
screen / menu.
3.
Press Next to scroll to System Maintenance, and press Select.
4.
Input the daily System Maintenance Code (obtained from ITE). To do so:
• Press [+] repeatedly to increment a digit.
• Press Next to move the cursor to the next digit.
5.
With the code entered, press Next to move the cursor to OK, and press Select.
6.
In the service menu, press Next to scroll to Calibration, and press Select.
2
ALCOLOCK V3 SERIES B-2