The general data transfer has the following format:
AAA YYYY-MM-DD hh:mm:ss+LF+CR (25 characters)
V
0000.0C PPP Exxx Ix+LF+CR
V
V
V
LF+CR
AAA:
address of storage position
YYYY-MM-DD: date (year, month, day)
hh:mm:ss:
time, (hour, minute, second, output in 24h-format)
0000.0:
temperature
C:
temperature in °C or °F
PPP:
transfered parameter (i.e.: MAX = Maximum value)
Exxx:
Emissivity in percent
Ix:
storage interval
The pyrometer always sends 10 data sets per second.
The settings of MIN, MAX and DIF do not change the data output.
In the AVG mode, the data sent to the RS-232 always represent the average
value.
1
at online-measuring "000"
2
instruments without data storage always filled out with zero
V
V
2
(22 characters)
(2 characters, empty line)
1
2
13