Direct Supply Panacea Air Element Mattress Manuel du propriétaire - Page 15
Parcourez en ligne ou téléchargez le pdf Manuel du propriétaire pour {nom_de_la_catégorie} Direct Supply Panacea Air Element Mattress. Direct Supply Panacea Air Element Mattress 20 pages. Pump and mattress system
Appendix A: EMC Information
Guidance and manufacturer's declaration – electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Emissions test
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
Immunity Test IEC 60601 Test Level
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power supply lines
transient/burst
IEC 61000-4-4
±1 kV for input/output lines
Surge
±1 kV line(s) to line(s)
IEC 61000-4-5
±2 kV line(s) to earth
interruptions and
<5% UT
voltage variations
(>95% dip in UT)
on power supply
for 0,5 cycles
input lines
40% UT
IEC 61000-4-11
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the A.C. mains voltage prior to application of the test level.
Compliance
Electromagnetic Environment – Guidance
Class A
This device is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
Not applicable
buildings used for domestic purposes.
Compliance
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
<5% UT
(>95% dip in UT)
for 0,5 cycles
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec.
3 A/m
3 A/m
Electromagnetic Environment –
Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions, it
is recommended that this device be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
1-800-634-7328
DirectSupply.com
15