Directions for Use

Directions for Use
1. P lace the Direct Supply Panacea Air Overlay on top of the mattress (not supplied)
with the hoses exiting at the foot of the bed.
2. Secure the overlay to the mattress by tucking the flaps located at the head and
footends under the mattress to prevent movement.
3. H ang the control unit on the end of the bed using the hangers, or place on a
stable horizontal surface.
4. E nsure that the air lines are not "kinked" or "twisted". Connect the overlay to the
control unit by pushing the connector into the air ports located on the side of the
pump. The hoses can be connected to either port on the pump unit.
5. E nsure that the electrical power source, if switched, is in an OFF position,
then plug in the power cord and switch to ON. The POWER ON lamp is
now illuminated.
6. S et the firmness level
for the overlay to fully inflate.
7. A fter preparing the overlay surface with appropriate linens (not included), place
the resident on the overlay.
8. D ecrease firmness
into the overlay for optimal pressure redistribution.
a) A common method for sizing an air overlay is to check to see if suitable
pressure is selected by sliding one hand between the air overlay and the air/
foam base under the resident's buttocks. Users should be able to feel the
space between their hand and the resident's buttocks with the acceptable
range being 25-40mm (1" to 1
AHCPR (Agency for Health Care Policy and Research).
b) A djust the firmness
immersion (if too much space during hand check, too firm) or clockwise for
less immersion (not enough space during hand check, too soft).
9. A n audio-visual alarm is sounded in the event of low pressure.
Audio alarm can be muted using alarm silence key.
CAUTION – Be careful not to puncture the overlay with needles or sharp
instruments. This may result in loss of integrity of the cover or internal air system.
*Firmness adjustment not available on Panacea Air Overlay with non-adjustable pump
8
*
to the highest setting and allow approximately 20 minutes
*
if necessary for resident comfort, support, and immersion
"). This hand check procedure is issued by
1
2
*
by turning the dial; counter clockwise for more