Direct Supply Panacea 1000 Manuel du propriétaire - Page 14
Parcourez en ligne ou téléchargez le pdf Manuel du propriétaire pour {nom_de_la_catégorie} Direct Supply Panacea 1000. Direct Supply Panacea 1000 20 pages. Adjustable-height low bed
Servicing
Troubleshooting
Symptom
No Power
Actuators operate,
but the wrong
movement occurs
One or more
actuators do
not function
Bed is not level
when operating
Hi/Lo functions
14
Possible Cause(s)
1. Bed is not plugged into
A/C power
2. Cut/damaged power cord
3. Loose plug for pendant. Cut/
damaged pendant cord
4. Power cord not connected to
control box
5. Thermal fuse in control box has
tripped
1. Cut/damaged pendant cord
2. Actuators are plugged into the
wrong ports on the control box
1. Loose or unplugged connection
of actuator cord to control box
2. Cut/damaged actuator cord
3. Defective actuator
4. Defective control box
5. Check for obstructions
1. Bed bumped a wall or
encountered other
obstruction causing
misalignment of actuators
2. Bed repeatedly being operated in
the mid-range loaded at one end
Action
1. Check plug to receptacle
2. Check power cord for damage
3. Check pendant cord
for damage
4. Check power cord connection to
control box
Inspect power cord for damage
5. The automatic resettable thermal
fuse will restore normal power
when a safe temperature is
reached. The time to reset may
be 10 to 15 minutes.
Ensure that the bed is operating
within the appropriate
environmental conditions, as
listed on page 7 of this manual.
Check for any obstructions that
may have hindered the bed's
ability to move freely. Remove
obstructions as necessary.
1. Inspect pendant cord
for damage
2. See control box socket
identification showing port/
socket locations
1. Check actuator plug to
control box
2. Inspect actuator cord
for damage
3. Replace actuator
4. Replace control box
5. Remove obstructions
1. Run bed to full up or full
down position after removal of
obstructions
2. Run bed to full up or full
down position