INSTALLATION
T
he UPS unit is designed for installation in a
protected environment. Some important
points to consider when positioning a unit for
operation:
CAUTION:
UPS SHALL NOT BE USED WITH
LIFE SUPPORTING EQUIPMENT AND SHALL BE
NON-PATIENT CONTACT EQUIPMENT.
CAUTION:
EXTERNAL BATTERIES ARE NOT
TO BE CONNECTED IF THE UNIT IS USED IN
THE PATIENT VICINITY.
A properly rated (preferably dedicated) outlet is
accessible for the six-foot power cord, supplied
with the unit. It is not recommended to modify the
supplied cord in any way nor should an extension
cord of any kind be used.
The cord paths in the system installation should
remain clear of foot traffic or anything else that
may disturb permanent connection.
The installation site should maintain an ambient air
temperature of less than 104
environment for the system remains cooler during
operation, there is less stress on the batteries and
the internal electronics.
The air inlets, vents and fan should not be
obstructed or blocked in any way. The more
breathing space the system has, the cooler it
operates.
The air should remain free from excessive dust
and chemical fumes.
Once a location has been selected and the unit is
installed, it is ready for operation.
Allow at least 8 hours, after the system is first
installed, to fully charge the internal battery to a
maximum state.
o
o
F (40
C). When the
Page 11
T H E C O N T I N U O U S P O W E R C O M P A N Y
ACCESSARY EQUIPMENT: Connected to analog
and digital interface must be certified according to
the respective IEC standards (i.e. IEC 60950 for
data processing equipment and IEC 60601-1 for
medical equipment). Furthermore all configuration
shall comply with the system standard IEC 60601-1.
Everybody who connects additional equipment to
the signal input part or signal output part configures
a medical system, and is therefore responsible that
the system complies with the requirements of IEC
60601-1. If in doubt, consult the technical services
department or your local representative.