BP filter (bandpass filter)
The bandpass filter passes only light of a certain
wavelength, blocking all other wavelengths.
BP filters are used for microscopy of fluorescent
images involving a specific dye in multiple-dye
specimens. (For example, in a double-dye
specimen of FITC and TRITC, the BA520-560 filter
enables microscopy of just the FITC fluorescent
image.)
However, BP filters will not indicate
self-fluorescence, if any (because the fluorescent
image in the above combination is green only). LP
filters are better suited for making fine distinctions
in self-fluorescence based on slight color
differences.

Replacing excitation and barrier filters

18.3
Excitation and barrier filters can be removed from
the filter cube and replaced. (Dichroic mirrors
cannot be dismounted from the filter cube.)
Excitation filters are screw-in filters.
Barrier filters are slide-in filters. Align the
projection on the barrier filter with the groove on
the filter cube and turn clockwise approximately 30
degrees to secure in place.
Individual Operations
18 Selecting Fluorescent Filters
33
Waveband
emitted by
FITC
Wavelength
Only the FITC fluorescent image is visible.
Turn about 30° to secure in place.
Excitation
filter
BA520-560 (BP type)
Waveband
emitted by
TRITC
Barrier filter