Nikon N-STORM Manuel de l'utilisateur - Page 9
Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Nikon N-STORM. Nikon N-STORM 11 pages. Super resolution microscope
Également pour Nikon N-STORM : Manuel d'utilisation (48 pages)
1. Open the settings window by pressing on the Settings button of the N-STORM module
and select the "Sequence Panel"
2. Acquisition should be in Image Data Acquisition Priority.
3. Set activation to Continuous.
1. Press options on the Settings window.
2. Remove the Z-stack option tick.
3. Camera options: 256x256 image and keep recommended settings OUT.Set image
acquisition parameters: by pressing on the "Edit" button of the N-STORM window ->
405/647: 10000 frame.