Armstrong Monitoring Corporation AMC-CD2 Manuel d'instruction - Page 9
Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Armstrong Monitoring Corporation AMC-CD2. Armstrong Monitoring Corporation AMC-CD2 12 pages. Portable carbon dioxide detection system, human occupancy detector .
AMC-CD2 Portable CO2 Detector
4 OPERATION
4.1 TURNING ON AND STABILIZING THE MONITOR
Turn on the unit with the power switch, as shown in Figure 3.1.2.1. The display will activate. The
sensor requires approximately 30 seconds to stabilize. The internal pump will turn on
immediately and run continuously.
Operating time with fully recharged batteries is approximately twelve (12) hours for the unit with
the LCD display and six (6) hours for the unit with the LED display. Power can be turned off to
conserve battery life.
4.2 TAKING AN AIR SAMPLE
The internal pump continuously draws air into the sensor chamber. To use the unit as a
concealed human occupant detector connect the sampling hose to the quick connect fitting (see
Figure 4.3.1.1) and use the wand fitted with the appropriate tip to obtain a sample. Ambient
atmospheric CO2 is within the range of 300 to 400 ppm but may range as high as 650 ppm
under certain environmental conditions. Any enclosed space with respiring humans will give an
elevated reading.
Warning:
Do NOT immerse the sampling wand in any liquid as it will damage the pump
mechanism. Turn the power off immediately if a liquid appears inside the
filter.
4.3 CHARGING THE MONITOR
The AMC-CD2 is powered with a sealed lead acid battery. The unit may be recharged at the
end of each operating session. The AC recharger / adapter plugs into the jack on the side panel
of the detector. ( see Figure 3.1.2.1 )
Full recharge is reached in approximately 12 to 16 hours maximum with the unit powered OFF.
The battery will increase to its full capacity after 25 charge / discharge cycles.
It is possible to power the unit directly from the AC recharger / adapter. The batteries will not be
charging while the unit is switched ON.
7