EMULATOR MODULE
MODEL # MOD-EMU-BWS
The Emulator module comes with a male
IEP connection (Infinite Expandability
Port). Prior to start-up, first ensure all the
toggle switches for the "Module Alarms"
Emulation Module
and "Lamp Alarm" are in the OFF (toggle
up) position. Plug the Emulator connector
Days
UV
into the IEP on the BLAKE/HO controller,
Solenoid Status
Closed
Open
then plug in the controller power cord.
Module Alarms
Sensor
Solenoid
4-20
Remote Alarm
Lamp
To Emulate a particular module failure,
toggle the corresponding switch to the
4-20 mA module
ON (toggle down) position. The alarm can
be cleared by returning the switch to its
MOD-EMU-BWS
OFF (toggle up) position. The device also
can emulate the UV intensity reported by
a UV sensor module. Turning the right "UV" knob in the clockwise direction will increase the
intensity, while turning it counterclockwise will decrease the intensity. Please note that to see
all the capabilities of the sensor, you will need to cycle the intensity through the range of 49
through 66 %. The "Days Remaining" cycle can also be adjusted by turning the left "Days" knob.