Operating Instructions
Operating Instructions
5-2
LED State
The LED will have a steady flash rate
(approximately 1 second ON and 1 second
OFF) when the basin is being heated up to its
programmed set-point temperature. When
Flashing
the basin temperature reaches 2°C below the
set-point temperature, the LED will change to
Steady ON, indicating that the control system
is now maintaining temperature within the
programmed temperature range.
On VP 2000 slide processors or heating modules with serial
numbers 32000 or greater, the following additional LED states
may be observed, representing two fault conditions of the
heating module:
Table 5.2:
Status LED Fault Conditions for Select Models
(Serial number 32000 and higher)
LED State
If a basin heater is stuck on when it should be
turned off, the LED will have a rapid flash rate
(approximately ¾ of a second ON and a ¼
Rapid
second OFF). If this occurs, the VP 2000
flashing
processor should be immediately powered
down and a Field Service Representative
contacted for assistance.
If a fault with the basin temperature sensor is
detected, the LED will have a slow flash rate
(approximately 2 seconds ON and 4 seconds
OFF). If this occurs, the basin heater will be
Slow
turned off automatically. If slow flashing is
flashing
observed, the VP2000 processor should be
powered down and a Field Service
Representative should be contacted for
assistance.