Calibration and pipe occlusion
A magnetite sample was used for factory calibration, This is a
cuvette 1cmx1cmx1.5cm packed with magnetite flakes. A force
measurement is recorded with the centre of a long side of the
cuvette 5mm from the centre of the probe measurement
surface (to simulate a pipe wall thickness). This value is taken
for the factory calibration gain and the software assumes that
this value corresponds to an occlusion of 75% in a tube of
40mm inner diameter and 5mm wall thickness.
By measuring the mass of the cuvette the density of the
magnetite flake sample was calculated and compared to the
density of solid magnetite. This gave a packing factor of 48%.
This factor is used in the display of occlusion for solid
magnetite. Calibration settings can be altered using the
<Calibrate> menu. The calibration standard supplied is
designed to mimic the above, measurement should be on the
black side without the label.
The offset has been set to be nominally 0 counts with the
probe held horizontally. In the unlikely event that the offset
needs to be adjusted then this can be carried by adjusting the
potentiometer. Gently peel back the calibration label on the
probe to access the potentiometer. Using a small screwdriver,
clockwise will take the offset negative, anticlockwise will take
the offset positive. Replace the calibration label after
adjustment. Note that the meter's range is nominally 0 to 2047
force units.
Page 8