case of breathing mixes for diving, the results of the analysis
may not be used as confirmation of the correct mix
composition. In these cases, the analyzer can only be used as
an aid for increasing the probability that such incorrect mix
composition will be detected before its use. The correctness of
the mix composition must be established by the technology
used to prepare it.
The correct preparation of breathing mixes for diving (NITROX
included) requires special knowledge and experience which is
possible to gain, for example, through an appropriate course. A
failure to apply such knowledge can lead to an incorrect mix,
even though the results of the analysis are seemingly correct.
Such a situation can also be the result of a failure to take into
account the influence of temperature and compressibility, an
insufficient blending of the mix and other factors.
Please remember that the analyzer can be broken or damaged
and that the oxygen sensor is by its nature subject to ageing
and thus its properties deteriorate.
Therefore always verify the measured data in another way, such
as by precise preparation of the mix, another analytical
method, etc. In addition, the measured value must be subject
to critical assessment using a common-sense approach and if
it differs from the allowable variation of mixing accuracy, it
must be considered invalid.