DJO Global 3045 Manuel - Page 3

Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} DJO Global 3045. DJO Global 3045 4 pages. Mimics ambulation and reduces dvt rates

3. Connect System
Connect cuffs to the tubing that is attached to the pump.
5. S Mode Operation
When powered on, the VenaFlow Elite System will default
to standard rapid inflation mode. If a slower inflation is
desired, select the S mode operation
Once selected, both cuffs will inflate simultaneously once
per minute.
Both cuffs will inflate and compress for a total of 10 seconds.
While in S mode the icon
of the graphical display. The device will remain in S mode
until user presses the S mode operation button again.
This button allows user to toggle between standard rapid
inflation mode and S mode.
Check Tubes Alarm
(Kinked or Disconnected Tube)
Intended Use/Indications
The VenaFlow Elite System is intended as a prophylaxis for deep vein thrombosis (DVT).
Contraindications
The VenaFlow Elite System should not be used by persons with known or suspected deep vein thrombosis, severe congestive heart failure, pulmonary edema, thrombophlebitis, severe arteriosclerosis, or active infection. Do
not use on extremities which are not sensitive to pain, where cuff will interfere with gangrene, on patients with vein ligation or recent skin grafts, or extreme deformity of the leg. Do not use the VenaFlow Elite System where
increased venous or lymphatic return is undesirable.
Warning and Precautions
• The VenaFlow Elite cuffs are designed for single patient use only.
• Risk of explosion if used in the presence of flammable anaesthetics.
• The equipment or system should not be used adjacent to or stacked with other equipment and, if adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
• Medical Electrical Equipment needs special precautions regarding EMC. Portable and Mobile RF Communications Equipment can be affected by other Medical Electrical Devices.
• Cuffs used in combination with warming devices may cause skin irritation. Regularly check for patient comfort, compliance and skin irritation.
• Special attention should be given to those patients with neuropathies or tissue viability problems (i.e. diabetes, arterial or venous insufficiencies).
• To prevent extremity compartment syndrome, special attention should be given to patients who are positioned in the supine lithotomy position for extended lengths of time. This includes patients with or without
compression cuffs.
• If you experience pain, swelling, sensation changes or any unusual reactions while using this device, consult your medical professional immediately.
button.
will appear in the top center
Call For Service Alarm
(Ship back to DJO for repair)
4. Standard Rapid Mode Operation
Press ON/OFF/RESET button.
Cuff inflation alternates between the two legs every 30
seconds.
Distal aircell inflates first and the proximal aircell follows.
Pressures settle at 45mmHg ± 15%. (Foot Cuff peak ~130
mmHg.)
After six seconds the cuff deflates.
Alarm will sound if the system is not functioning properly.
6. Locking in Standard Rapid
or S Mode
To lock in standard rapid mode, make sure the device is in
standard rapid mode, then press and hold the S mode
button for 5 seconds. Do the same to unlock it.
To lock in S mode, make sure the device is in S mode and
press and hold the S mode button
the same to unlock it.
Once the system is locked in either standard or S mode, the
lock symbol
will appear on the screen.
Alarm Reset
To reset any alarm, press the center
ON/OFF/RESET
take steps to correct the alarm if
necessary.
Note: Pressing ON/OFF/RESET
following alarm only
The device will continue to run and will
not shut down. Alarm will continue at
next cycle if not corrected.
www.DJOglobal.eu
for 5 seconds. Do
button and
resets
the alarm.