6.1.1. Analog / Digital Gain Settings
ARTCAM-990SWIR-TEC-CL is compatible with analog and digital gain.
Setting value of gain on software can be set from 0 ~ 420, and the gain will change from 0.0dB ~ 42.0dB
accordingly. Gain can be calculated by the following formula:
Gain [dB] = value / 10.0[dB]
Digital gain will activate when setting value of gain > 180.
The graph below shows the relationship between setting value of gain and its magnification[dB].
6.1.2.
Shutter speed calculation
ARTCAM-991SWIR-TEC-CL allows for shutter speed control.
Shutter speed value can be calculated by the following formula:
Shutter speed [us] = value*12.96[us]
- 9 -