ARTROMOT K3 Manuel d'entretien - Page 2
Parcourez en ligne ou téléchargez le pdf Manuel d'entretien pour {nom_de_la_catégorie} ARTROMOT K3. ARTROMOT K3 13 pages.
Également pour ARTROMOT K3 : Manuel d'entretien (12 pages)
Table of contents
1. History
9. Explosion drawing part 2
10. Checklist for safety- and function test
1. Historie History
Revision Date
Name
1
17.09.2001 M. Nunnenmacher Initial Release
2
19.09.2001 M. Nunnenmacher Pos. 4.8
3
15.10.2001 M. Nunnenmacher Pos. 1.9, 1.15,
4
6.11.2001
M. Nunnenmacher Update
5
15.01.2002 M. Nunnenmacher Update Pos.
6
06.02.2002 M. Nunnenmacher Update Pos.
7
27.03.2002 M. Nunnenmacher Update Pos. 1.1
8
21.10.2002 S. Herr
9
02.06.2003 S. Herr
10
05.04.2004 S. Herr
11
11.07.2006
S. Herr
2. Purpose
2
This service manual is to perform some repairs on
ARTROMOT®-K3 products. Repair and maintenance work
2
may only be carried out by authorised persons, as other-
wise all warranty services and liabilities shall be void.
2
It is not allowed to utilize any other spare part which is
not mentioned in chapter 6.
4
4
5
3. General
9
10
3.1 Electronic and cable
11
Electronic devices as hand-held programming unit, motor
electronics and spiral cable are not interchangeable with
12
ARTROMOT®-K3 serial number < 3000.
Make sure that the characteristic values of your power
supply correspond to the voltage and frequency data
indicated on the power adapter.
Change
Error possibilities, showed by the hand-held pro-
gramming unit:
DEVICE NEEDS
REFERENCE RUN: The reference run was interrupted
or the electronics were replaced.
-> Perform a reference run. See chapter 7.4
1.16
POTENTIOMETER: Potentiometer shows incorrect value
-> Check plug connection, replace potentiometer if
reference run,
persistent
new packaging
POTENTIOM. CABLE: Spiral cable for knee case
defective
3.13, 3.45, 4.3,
-> Replace spiral cable for knee case (Pos. 5.9),
4.5, Safetytest
check plug connection
3.12 Labeled
MOTOR DRIVER: Motor electronics defective
casting
-> Replace motor electronics (see chapter 7.3) if
persistent
OVER CURRENT: Motor power to high
-> Replace motor electronics (see chapter 7.3) if
Update for all
persistent
electronic parts,
explosion
UNWANTED BUSY: Motor control unit is not answering
drawings and
-> Hand-held programming unit, motor electronics or
reference run
spiral cable of the hand-held programming unit defective
Update Pos.
CPM MEMORY RAM: Memory fault in the motor control
unit
-> Replace motor electronics (see chapter 7.3) if
Update Pos.1.6,
persistent
CPM MEMORY ROM: Memory fault in the motor control
Update
unit
chapter 4, 8, 9
-> Replace motor electronics (see chapter 7.3) if
Pos. 1.25 -1.28,
persistent
Pos. 4.20 - 4.26
2