1-CUBE GMD Manuel de l'utilisateur - Page 13

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} 1-CUBE GMD. 1-CUBE GMD 14 pages. Measuring device of co2 content in fermenting tanks and keg casks

5.2.4 R
EADING RECORDED VALUES OF
The measured values of CO2 are kept into the device non-volatile memory. The capacity of memory
is 450 measured values.
1) Press the key ON to switch on the device. The text „1-CUBE" with the program version appears
on the display for a moment. Then the following text appears:
„ENT.meas." | „+ view."
2) Press the key plus (+). There is text „memory" on the first line and the measured values of CO
in g/l appear in turn from the oldest value to the newest ones on the second line.
after the last saved value the numbers start to appear 655.35 which indicates blank memory
values. Press the key ENTER to stop scanning memory. Then the following text appears
„ENT.meas." | „+ view."
Note: Once memory is full, then it will be automatically deleted.
Then the following text appears:
„memory" | „deleting"
All the saved values are overwritten with number 655.35.
5.2.5 D
ELETING THE CONTENT OF THE MEMORY
The measured values can be deleted from memory at any time. The memory capacity will be
completely recovered (450 values) and newly measured values will be saved in sequence.
1) Press the key ON to switch on the device. The text „1-CUBE" with the program version appears
on the display for a moment. Then the following text appears:
„ENT.meas." | „+ view."
2) Press the key minus (-).Then the following text appears: „memory" | „deleting"
All the saved values are overwritten with number 655.35.
6.0 CO
TRANSFERRING MEASURED VALUES FROM MEMORY TO
2
(
COMPUTER
TYPE
Only type GMDK has the functionality of data transfer with a communication cable to PC. Before the
first data transfer from the device to the PC, first install program DataViewer from our supplied
CD.
1) Connect the shut-off device GMDK with the help of cable FTDI USB-RS485-WE-1800-BT
(supplied with the device version GMDK) with USB port of your PC. Wait until the driver is fully
installed and ready. This occurs the first time you plug the USB adapter into a USB port of your
computer.
CO
FROM MEMORY
2
GMDK
)
ONLY
2
PC
13