1-CUBE SU 6.2. Manuel de l'utilisateur - Page 10

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} 1-CUBE SU 6.2.. 1-CUBE SU 6.2. 13 pages. Water thermostat

Calibration of the lower temperature 5 degrees Celsius (point 1) :
First check if the temperature in bath is about 5 degrees Celsius and if it is not use the method Thermostat to reach it.
Once temperature of 5 degrees Celsius is reached use the following procedure:
1) Press the keys OFF and ENTER at the same time
2) There is text „dig value of tem" on the first line and the value on the second line. Write down the digital value.
3) There is text „point 1 - digital" on the first line and the value on the second line. Overwrite the value with the
help of keys
+ (increase ) or – (decrease) to put the digital value that you have written down before. Press the
key „OFF" to set the required value.
4) There is text „point 1 - temp" on the first line now and the value on the second line. Overwrite the value with the
help of keys
+ (increase ) or – (decrease) to put the value that you have measured before with the etalon in
the water of the water bath. Press the key „OFF" to set the required value.
5) There is text „point 2 - digital" on the first line now and the value on the second line. Do not set the point 2 at this
moment therefore press the key „OFF".
6) There is text „point 2 - temp" on the first line now and the value on the second line. Do not set the point 2 at this
moment therefore press the key „OFF".
7) There is text „HYSTER1_0= " on the first line now and the value on the second line. Do not set HYSTER1_0= at
this moment therefore press the key „OFF".
8)
There is text „HYSTER1_100= " on the first line now and the value on the second line. Do not set HYSTER1_100=
at this moment therefore press the key „OFF".
9) There is text „HYSTER2_0= " on the first line now and the value on the second line. Do not set HYSTER2_0= at
this moment therefore press the key „OFF".
10) There is text „HYSTER2_100= " on the first line now and the value on the second line. Do not set HYSTER2_100=
at this moment therefore press the key „OFF".
11) There is text „HYSTER3_0= " on the first line now and the value on the second line. Do not set HYSTER3_0= at
this moment therefore press the key „OFF".
12) There is text „HYSTER3_100= " on the first line now and the value on the second line. Do not set HYSTER3_100=
at this moment therefore press the key „OFF".
The lower temperature (point 1) is calibrated now.
10