Etac Molift Air Series Manuel de l'utilisateur - Page 14

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} Etac Molift Air Series. Etac Molift Air Series 15 pages.

Etac Molift Air Series Manuel de l'utilisateur
Molift Air
Cleaning and disinfection
Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the lifter. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products. Check
emergency stop and emergency lowering after
cleaning. The lift should not be exposed to run-
ning water.
Make sure not to damage or
remove labels when cleaning.
Periodic Inspection Scope
Periodic inspection is a visual examination (par-
ticularly of the hoist's load bearing structure and
lifting mechanism with attachments, brakes, con-
trols, safety devices and person-support devices)
according to Periodic Inspection Report for Molift
Air.
Periodic Inspection shall be performed at least
once a year or more frequently if required by local
requirements. The inspection must be performed
by service personnel authorized by Etac. Contact
Etac for training and authorization or recommen-
dation of an approved service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report for
Molift Air. The reports should be retained by the
person(s) responsible for servicing the hoist. If
the inspection reveals defects and damages, the
owner shall be notified and a copy of the report
should be sent to Etac.
In the event of danger to safety,
the hoist shall immediately be
taken out of service and marked
clearly with "out of order" and
shall not be used until the hoist is
repaired

Maintenance

Reconditioning

Follow cleaning and installation procedure, com-
plete periodic inspection and use checklist after
installation to recondition the lifter.

Recycling

Refer to "Recycling instructions" for how to
properly dispose of product.

Spare Parts

A list of spare parts is available on request.

Periodic Inspection

When periodic inspection is completed the
inspector shall mark the hoist with a sticker on
the control label showing the date when periodic
inspection is performed. The control label can be
found on the inside of the Side cover, and this
will then indicate when next service should be
performed.
CONTROL
12 Month
13
Mark label
with month
and year of
inspection