OperatiOn:

nOte: follow the initial setup procedure before initial operation.
1.
Press the 'OPEN / STOP' button to unlock the lid and then open the lid.
2.
Insert cushions (if needed) into the tube carriers (holders) for the tube size you are
using. Refer to 'Tube Holder Configurations' for assistance.
3.
Place the test tube samples into the tube carriers (holders). Be sure to follow the
rules for balanced loads.
4.
Close the lid and turn the lid knob clockwise to its complete stop position. The
'LATCHED' indicator light should illuminate to indicate that the latch is closed
properly. If the lid knob is not completely latched, the 'LATCHED' indicator light will not
illuminate and the centrifuge cannot be operated.
5.
Set the desired speed and run time using the appropriate up and down arrow buttons.
6.
If desired, adjust the braking level, acceleration rate, countdown delay, imbalance
sensitivity and audible indication. Turn to page 11 for a description of these features.
7.
Begin the run by pressing the 'START' button on the control panel.
8.
The centrifuge should begin to spin. The 'RUNNING' indicator light should illuminate.
if a prObleM is fOunD DurinG a spin that reQuires the centrifuGe tO shut
DOWn, press the 'Open / stOp' buttOn iMMeDiatelY !
9.
After time has elapsed, the 'RUNNING' indicator light will extinguish and the rotor will
decelerate to a complete stop.
11.
The 'UNLOCKED' indicator light will illuminate and the locking mechanism will
disengage, allowing entry into
(after 60 seconds) simply press the 'OPEN / STOP' button.
12.
Turn the lid knob counterclockwise and open the lid.
13.
Remove the samples.
14.
The centrifuge may be used immediately.

Rotor Configurations:

The next section describes proper balancing and tube loading. Use the appropriate section for the
rotor installed in your centrifuge. Improper loading can result in damage to the centrifuge or test
tube samples.
12 Place combo Fixed-Angle STAT Rotor (Standard):
This 12 place rotor is designed to hold six (6) 17x100 mm and six (6) 17x75 mm tubes.
The supplied adapters allow for operation with 13x100mm and 13x75mm tubes.
Your centrifuge must contain a balanced load in order to work properly. To ensure
that the load is balanced, keep these rules in mind when inserting test tube samples.
1. Opposing tube locations must be empty or loaded with equally weighted samples.
2. If an odd number of samples is to be spun, use a water-filled tube to mate with the
unpaired one.

Default user settinGs:

nOte: Your centrifuge comes pre-set with the following settings:
Braking Rate: 0
Acceleration Rate: 9
Countdown Delay: ON
Imbalance Detection: LOW
Audible Setting: 4
the rotor chamber. If the automatic unlock times out
(See page 8/9 for a description of the user settings)