PROBLEM: Table height adjustment will not work properly.
SOLUTION: Make sure the table is connected to a working electrical outlet. Inspect all power cords
for cuts or breaks that would interfere with the electrical connection. Check to make certain foot pedal
cord is properly and securely plugged into motor. If after checking all these areas, the table will not
operate, or operates improperly, call a technician, or call Clinton Industries at 800-441-9131 for
assistance. NEVER USE THE TABLE IF IT IS NOT WORKING PROPERLY.
REMOVAL OF THE MOTOR FOR SERVICE
The motor assembly is located at the foot-end of the table on the base. Next, detach the motor
assembly. If possible, raise the table to its highest position. Raising the table will make it easier to remove
the motor mount bolts. Using help, lower the table onto its side onto a protective surface
such as a blanket or other material that will not harm the upholstered top. Remove the foot pedal
controller by unplugging the DIN plug on the side of the motor. Carefully remove the upper and
lower motor mount bolts to free the motor assembly.
For assistance with other problems not covered in this guide, or questions about your
Clinton Power Table, please call Clinton Industries at 800-441-9131.
Upper Motor Mount
Motor Assembly
Lower Motor Mount
Foot Pedal DIN Plug
11