• Do not get fluids inside the meter through the test strip port, data port or battery compartment.
• Never immerse the meter or hold it under running water because it will damage the meter.
• Two disposable wipes will be needed for each cleaning and disinfecting procedure; one wipe for cleaning and
a second wipe for disinfecting.
Cleaning
W ear appropriate protective gear such as
disposable gloves.
O pen the cap of the Clorox
container, pull out 1 towelette and close the cap.
W ipe the entire surface of the meter 3 times
horizontally and 3 times vertically using one
towelette to clean blood and other body fluids.
D ispose of the used towelette in a trash bin.
NOTE: No actual drying of the meter is necessary before starting the disinfecting procedure.
Disinfecting
(The meter should be cleaned prior to disinfection.)
P ull out 1 new towelette and wipe the entire surface
of the meter 3 times horizontally and 3 times
vertically to remove blood-borne pathogens.
Dispose of the used towelette in a trash bin.
Allow exteriors to remain wet for 1 minute, then
wipe the meter using a dry cloth.
A fter disinfection, the user's gloves should be
removed andthrown away. Wash hands before
proceeding to the next patient.
If any of the following deterioration signs appear after cleaning or disinfecting, please stop using the system and
contact Customer Service at 800.818.8877, 24 hours a day, 7 days a week, USA and Canada.
• When the inscriptions on the exterior of the meter have been removed.
• When the color of the meter has changed or faded.
• When cracks or roughness develop on the meter.
• When a part of the segment on the meter display does not flash.
• http://www.cms.gov/Regulations-and-Guidance/Guidance/Transmittals/index.html
• http://www.cdc.gov/injectionsafety/blood-glucose-monitoring.html
• www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
• http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html
Minneapolis, MN 55439 • 800.818.8877 • www.arkrayusa.com
CLEANING AND DISINFECTING PROCEDURES
®
Germicidal Wipes
More information is available at:
Assure
®
Prism Multi Blood Glucose Meter
• For multiple patients / professional use.
• Use with Assure Prism Multi Blood Glucose Test Strip.
• For testing outside of the body (in vitro diagnostic use only).
• Model: GM505RA
3V, 11 mA (CR2032 X 2)
Manufactured for
Minneapolis, MN 55439 USA
800.566.8558
www.arkrayusa.com
Made in Korea
Front
Back
1
minute
1
minute
Contact Customer Service
at 800.566.8558, 24 hours
a day, 7 days a week, USA
and Canada.
P/N 5117-00 REV 07/14
Both sides
Assure
®
Prism Multi Blood Glucose Meter
• For multiple patients / professional use.
• Use with Assure Prism Multi Blood Glucose Test Strip.
• For testing outside of the body (in vitro diagnostic use only).
• Model: GM505RA
3V, 11 mA (CR2032 X 2)
Manufactured for
at 800.566.8558, 24 hours
Contact Customer Service
a day, 7 days a week, USA
Minneapolis, MN 55439 USA
and Canada.
800.566.8558
www.arkrayusa.com
Made in Korea
P/N 5117-00 REV 07/14
dry cloth
©ARKRAY USA, Inc. All rights reserved.
6579-01 Rev 03/15