HOW TO OPERATE YOUR CLIPSULIN
1. SETUP
3. USING THE CLIPSULIN
Install a new battery.
Remove the cap of your insulin
pen and install the needle. Click the
·
Battery type: CR2032.
power button to turn on the CLIPSULIN
·
Refer to "Battery Replacement"
section.
Clip the CLIPSULIN
onto your insulin
®
pen in a way that allows you to see the
LCD screen next to the unit window of
your insulin pen.
·
"00" will be displayed on the CLIPSULIN
·
Attach the CLIPSULIN
®
near the dosage
screen.
window.
Select your injection dose on the
·
Ensure that your CLIPSULIN
®
is tightly
insulin pen.
fitted onto your insulin pen.
CLIP
·
Ensure that your hands DO NOT come
into contact with the CLIPSULIN
during dose selection.
Verify that the number of units
displayed on the CLIPSULIN
same as the selected dose on the
insulin pen.
If the numbers are not the same, use
the " + " or " - " button to adjust
manually the units on the CLIPSULIN
2. BEFORE FIRST USE
Go to www.diabnext.com/app and
·
You can increase the number gradually
install the DIABNEXT
APP on your
®
by holding down the "+" (plus) button,
mobile device.
or decrease the number gradually by
holding down the "-" (minus) button.
Confirm the two numbers are the same,
then click the blue button (Power) to
validate and send your dosage data.
DIABNEXT
Turn on the Bluetooth function of your
mobile device and open the DIABNEXT
®
APP.
Enter the menu, select "My Tools" and
click "Add Tools".
·
The green light and the LCD will blink
Click "Insulin Pens" and select your
to signal that the data has been
insulin pen type.
recorded.
·
The data will not be recorded if the
Follow the instructions on the APP.
green light and the LCD have not
(Press and hold down the power button
blinked.
of CLIPSULIN
®
, until you see the double
·
Your data will be automatically
dash "- -" blinking on the LCD screen,
transmitted to your mobile APP
to pair with the APP.)
after injections. (See the " ★ " part.)
·
If you do not want your data to be
transferred automatically, please refer
to the "Data Transmission" section.
Perform your injection in your normal
manner.
·
It will take a few seconds to connect.
·
"- -" will be displayed on the screen
·
The device will automatically shut
of CLIPSULIN
.
®
down after 10 seconds of inactivity.
DATA TRANSMISSION
®
BLUETOOTH
®
You can use the Bluetooth transmitting
function with your mobile phone if you
.
have iOS / Android with BLE 4.0 or
®
higher.
Enable the Bluetooth function on
your mobile device.
®
Open the DIABNEXT
®
DIABNEXT
Follow the instructions on the APP.
®
is the
2
4
·
Press and hold down the power button
of CLIPSULIN
, until you see the double
®
dash "- -" blinking on the LCD screen,
to pair with the APP.
·
"- -" will be displayed on the screen of
CLIPSULIN
®
.
®
.
+1 UNIT
Your stored records will be transferred
2
automatically, and can be viewed on
4
your mobile device.
-1 UNIT
If you see the "Device error, please
contact customer service" message
2
2 2 2 2 2 2 2 2 2
pop up, please refer to the
4
4 4
"Troubleshooting" section.
The dose, date, and time of injection will be automatically stored in your CLIPSULIN
The CLIPSULIN
®
will automatically transmit the stored injection data under these conditions:
1. You have paired your CLIPSULIN
2. The Bluetooth/NFC function is enabled on your smartphone or tablet.
3. Your smartphone or tablet is nearby with the DIABNEXT
your CLIPSULIN
®
Your CLIPSULIN
can store up to 200 records, but the NFC connection can only transmit the
®
latest 50 records in the memory. Transfer your records regularly to your smartphone
or tablet to keep track of your injections.
NFC
You can use the NFC transmitting function
with your mobile phone if you have:
- iPhone 7 or newer, and have the iOS 11
system updated or newer.
- Android phone with NFC function.
Enable the NFC (Near Field
Communication) function on your
mobile device.
Note: For iPhone users, you may
skip this step.
APP.
Open the DIABNEXT
APP.
®
®
DIABNEXT
Follow the instructions on the APP.
·
Press and hold down the power button
of CLIPSULIN
, until you see the double
®
dash "- -" blinking on the LCD screen,
to pair with the APP.
·
"- -" will be displayed on the screen of
CLIPSULIN
.
®
Your stored records will be transferred
automatically, and can be viewed
on your mobile device.
NFC signal
emitting point
·
The NFC transmitter of the CLIPSULIN
is near the power button. Please
confirm the NFC transmitter location
of your smartphone/tablet, and be
sure to have your smartphone/tablet
close to your CLIPSULIN
®
for proper
pairing/syncing.
If you see the "Device error, please
contact customer service" message
pop up, please refer to the
"Troubleshooting" section.
Only the latest 50 records can be read
out via NFC connection. Transfer your
records at least once per week to keep
track of your injections.
®
with the DIABNEXT
®
APP before.
®
APP open amd connected to
.
BATTERY
BATTERY REPLACEMENT
Transfer your stored records before
replacing the battery.
Open the battery compartment.
Replace the old battery with a new
one.
Battery
Push the compartment back into
the CLIPSULIN
to close it.
®
Click the power button to check the
status of CLIPSULIN
.
®
Sync your device with the DIABNEXT
APP before using it again, to ensure
that the timestamps of the records
will be correct.
FCC STATEMENT
This device complies with part 15 of the
FCC Rules. Operation is subject to the
following two conditions:
1. This device may not cause harmful
interference, and
2. This device must accept any interference
received, including interference that
may cause undesired operation.
NOTE:
®
This equipment has been tested and found
to comply with the limits for a Class B
digital device, pursuant to part 15 of the
FCC Rules.
These limits are designed to provide reasonable
protection against harmful interference in a
residential installation.
This equipment generates, uses and can
radiate radio frequency energy and, if not
installed and used in accordance with the
instructions, may cause harmful interference
to radio communications. However, there
is no guarantee that interference will not
occur in a particular installation.
If this equipment does cause harmful
interference to radio or television reception,
which can be determined by turning the
equipment off and on, the user is encouraged
to try to correct the interference by one or
more of the following measures:
· Reorient or relocate the receiving antenna.
®
.
· Increase the separation between the
equipment and receiver.
· Connect the equipment into an outlet on
a circuit different from that to which the
receiver is connected.
· Consult the dealer or an experienced
radio/TV technician for help.
Changes or modifications not expressly
approved by the party responsible for
compliance could void the user's authority
to operate the equipment.
TROUBLESHOOTING
If there is no LCD display and/or LED signal light
feedback of CLIPSULIN
®
, please follow the steps below:
1.
Check if there is any physical damage to the
CLIPSULIN
®
. If yes, stop using the device and
replace it with a new one. If no, check the LCD
and LED functions.
2.
If only the LCD display OR the LED signal light
is not functioning, stop using the CLIPSULIN
®
and
replace it with a new one.
3.
If both the LCD and LED are not functioning
properly, replace the battery and check both
functions again.
*
Caution: After replacing the battery, before using
the CLIPSULIN
®
with your insulin pen, sync the
CLIPSULIN
with the DIABNEXT
APP to adjust its time.
®
®
4.
If the problem persists, go to customer service.
(www.diabnext.com/Q&A)
If the CLIPSULIN
cannot connect with the
®
DIABNEXT
APP, please follow the steps below:
®
1.
Check if there is any physical damage to the
CLIPSULIN
®
. If yes, stop using the device and
replace it with a new one.
2.
If no, reboot the Bluetooth/NFC function on the
mobile device, and then try to connect with the
CLIPSULIN
again.
®
3.
If rebooting the Bluetooth/NFC function does not
resolve the issue, reinstall the DIABNEXT
®
APP,
then try to connect with the CLIPSULIN
again.
®
4.
If the problem persists, go to customer service.
(www.diabnext.com/Q&A)
®
If the injection data cannot be transferred to the
DIABNEXT
APP, please follow the steps below:
®
1.
Check if there is any physical damage to the
CLIPSULIN
. If yes, stop using the device and
®
replace it with a new one.
2.
If no, reboot the Bluetooth/NFC function on the
mobile device, and then try to connect with the
CLIPSULIN
®
again.
3.
If rebooting the Bluetooth/NFC function does not
resolve the issue, reinstall the DIABNEXT
APP,
®
then try to connect with the CLIPSULIN
®
again.
4.
If the problem persists, go to customer service.
(www.diabnext.com/Q&A)
If you see the "Device error, please contact customer
service" message pop up in your APP, please follow
the steps below:
1.
When this message pops up, it means that
some records in the memory might be
damaged.
2.
Please confirm the message and go to customer
service.
(www.diabnext.com/Q&A)