Declaration – electromagnetic immunity
The AS-1001D suction pump is intended for use in the electromagnetic environment specified below. The customer or the user of the
AS-1001D suction pump should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV for input/output lines
IEC 61000-4-4
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
Voltage dips, short
< 5 % UT
interruptions and
(> 95 % dip in UT) for 0.5 cycle
voltage variations
40 % UT
on power supply
(60 % dip in UT) for 5 cycles
input lines
70 % UT
IEC 61000-4-11
(30 % dip in UT) for 25 cycles
< 5 % UT
(> 95 % dip in UT) for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Declaration – electromagnetic emissions
The AS-1001D suction pump is intended for use in the electromagnetic environment specified below. The customer or the user of the
AS-1001D suction pump should assure that it is used in such an environment.
Emissions test
Compliance
CE emissions
Group 1
CISPR11
RE emissions
Class B
CISPR11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
Complies
Flicker emissions
IEC 61000-3-3
Copyright © 2019 ANEST IWATA SPARMAX Co., Ltd. All rights reserved.
Compliance level
3V, 10V
3V/m, 10V/m
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV differential mode
±2 kV common mode
< 5 % UT
(> 95 % dip in UT) for 0.5 cycle
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
< 5 % UT
(> 95 % dip in UT) for 5 sec
3 A/m
Electromagnetic environment - guidance
The AS-1001D suction pump uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The AS-1001D suction pump is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
11 / 12
Electromagnetic environment - guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the EQUIPMENT or SYSTEM including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter. Interference may occur in the
vicinity of equipment marked with the following
symbol.
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
If the user of the EQUIPMENT or SYSTEM requires
continued operation during power mains
interruptions, it is recommended that the
EQUIPMENT or SYSTEM be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Cliq_AS-1001D Manual (230V50Hz_20190125.1)
3510406D