OPERATION:
After the unit has been set up and you have
made yourself familiar with the operation func-
tions, you are ready for operation as follows:
1. Turn the power switch on the lower left
front of the console to the 'ON' position.
The front panel display will light.
2. Attach a sterile handpiece to the motor
and set the Ratio Selector to match the
gear ratio of the handpiece being used
(i.e. 1:8 for a 1:8 handpiece, or 1:16 for a
1:16 handpiece).
3. Adjust the up/down arrows on the RPM
display to the desired operating speed on
the RPM readout.
NOTE: The RPM display indicates true
instrument speed when the Ratio Selector
is in the appropriate position. For display
accuracy, make sure the Ratio Selector
matches the actual handpiece ratio.
4. Adjust the 'Torque Control' to the desired
operating level. Begin at minimum levels
when using endodontic rotary instruments.
5. Depress the foot control to activate motor
and begin operation.
5.