Ashcroft EE Manuel d'instructions pour l'installation et l'entretien - Page 13

Parcourez en ligne ou téléchargez le pdf Manuel d'instructions pour l'installation et l'entretien pour {nom_de_la_catégorie} Ashcroft EE. Ashcroft EE 19 pages. Bimetal thermometer

Ashcroft EE Manuel d'instructions pour l'installation et l'entretien

10 Faults

All defective or faulty equipment must be taken out of service. Defective or faulty devices should be handed over to the
manufacturer's service department immediately. Under no circumstances repair attempts should be made on site.
Device safety can no longer be guaranteed.
Contact details see chap. 1.6
Fault
Jerky or random pointer movement
Bent or loose pointer
Cracked window
Leakage with filled device
Case damage
Signs of leakage of the measuring
system (discoloration of the dial or
the filling liquid).

10.1 Behaviour after rectifying the fault

See chapter 8 Assembly and Installation
11 Dismantling & disposal

11.1 Disassembly

During maintenance work on an unit installed without thermowell, the lines must be depressurized and the system
must be secured against being switched on again.
Dismantle the measuring device using a suitable tool.
Residual media in and on dismantled measuring instruments can endanger people, the environment
and equipment. Sufficient precautionary measures must be taken. If necessary, the devices must be
cleaned thoroughly (see notes in the safety data sheets).

11.2 Disposal

At the end of the product life cycle, do not dispose of this product with normal household waste. Take this
product to a collection point or a specialist disposal company for recycling of the components
Possible causes
Bended stem, bended every angle
connection
Damage with transport or external
rough impact
External influences
High solar radiation, operation
outside the device specification
Improper handling
Process media not compatible with
the material used for the thermowell,
installation location with excessive
external influences
page 13 of 19
Possible measures
Check for mounting errors and
installation location.
Replace device if necessary
Repair by the manufacturer
Repair by the manufacturer
Repair by the manufacturer
Check installation location,
Device must be replaced
Improper use,
Replace device
Adapt thermowell material to
process media