EG&G ORTEC 427 Manuel d'instruction - Page 17
Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} EG&G ORTEC 427. EG&G ORTEC 427 20 pages. Delay amplifier
F
6-2
(3) Connect the bipolar output of the 410 to the INPUT of the 427
through RG-62/U coble and terminate the cable at the input of
the 427 with a 100-ohm terminator. Ensure that all DELAY
switches are in the OUT position.
(4) Vary the amplitude of the 427 OUTPUT by adjusting the 419
Pulser; there should be no distortion of the signal as it is varied
from 0 to 10 volts.
(5) Raise the output amplitude of the 419 until the 427 saturates; the
saturation level should not be less than 1 1 volts.
(6) Adjust the 419 for a 10-volt output signal from the 410 and leave
it at this setting for the remainder of the test.
(7) The output of the 427 should be 10 volts and should come promptly
with respect to the input signal.
(8) Switch each individual delay line into the circuit and ensure that
the 427 output remains at 10 volts and is delayed the appropriate
amount.
(9) Switch al l delay lines into the circuit. The 427 output should be
approximately 11 volts and delayed 4.75 microseconds from the
input signal.
6.2 Changing Amplifier Gain
R22
The gain of the output coble driver loop is given by the ratio of
gain of this loop can be changed as much as 10% by increasing or decreasing the
value of Rll and R12 without any adverse effect. Since these resistors are used to
terminate the delay lines, it is not recommended that their values be changed by more
than 10%.
6.3 Suggestions for Troubleshooting
In situations where the 427 is suspected of malfunction, it is essential to
verify such malfunction in terms of simple pulse generator impulses at the input
and output. In consideration of this, the 427 must be disconnected from its
position in any system, and routine diagnostic analysis be performed with a test
pulse generator and an osci l loscope. It is imperative that testing not be per
formed with a source and detector until the Delay Amplifier performs satisfact
orily with the test pulse generator.
The testing instructions in Section 6. 1 of this manual, and the circuit descriptions
in Section 5 should provide assistance in locating the region of trouble and repairing