CELUTION PROCESSING
STEP 3: CELASE ADDITION
DO NOT ADD CELASE UNTIL PROMPTED
BY THE DEVICE!
15
A.
Aspirate 5 mL of LR into a luer-lock
syringe
B.
Reconstitute Celase using the 5 mL
LR solution and thoroughly mix
C.
Aspirate the Celase amount shown
on the device promp into a syringe
D.
Add the Celase to port "2" and cap
a er addition
STEP 4: CELL OUTPUT REMOVAL
Manufacturer:
Cytori Therapeutics, Inc.
3020 Callan Road
San Diego, CA 92121, USA
Tel: +1.858.458.0900
European Authorized
Representative:
RSQR Ltd.
Ludgate House
107 Fleet Street
London EC4A 2AB, UK
16
CUSTOMER SUPPORT
USA: +1.858.875.5245
EU: +44.1244.360.426
A.
Attach wrench as shown
Rotate clockwise
B.
C.
Remove the luer cap on the
processing chamber
D.
Using a spinal needle and
a 10 mL syringe, slowly
reconstitute cell output,
then aspirate into syringe
E.
Repeat 16A-16D on the
opposite side. Total output
volume is approximately
5 mL
CELUTION PROCESSING
IS FINISHED
T026-USA-EN_A