Ambu 620001000US Petunjuk Penggunaan Manual - Halaman 15
Jelajahi secara online atau unduh pdf Petunjuk Penggunaan Manual untuk Peralatan Medis Ambu 620001000US. Ambu 620001000US 17 halaman.
Sterilisation
Method of sterilisation
Operating environment
Temperature [°C, (°F)]
Relative humidity [%]
Atmospheric pressure [kPa]
Altitude [m]
Biocompatibility
Please be aware that the bending angle can be affected if the insertion cord is not kept straight.
1
There is no guarantee that endotherapy instruments selected solely using this minimum instrument channel
2
width will be compatible in combination.
Storage under higher temperatures may impact shelf life.
3
6. Troubleshooting
If problems occur with the system, please use this trouble shooting guide to identify the cause
and correct the error.
Problem
No live image on
the screen but User
Interface is present
on the display or
the image shown
is frozen.
aScope 5 Broncho 2.7/1.2 and 4.2/2.2
ETO
aScope 5 Broncho 2.7/1.2 and 4.2/2.2
10 – 40 (50 – 104)
30 – 85
80 – 106
≤ 2000
aScope 5 Broncho is biocompatible
Possible cause
The endoscope is
not connected to the
displaying unit.
The displaying unit
and the endoscope
have communication
problems.
The endoscope
is damaged.
A recorded image is
shown in the yellow file
management tab.
Recommended action
Connect the endoscope to the green
port on the displaying unit.
1. Reconnect aScope 5 Broncho by
unplugging and reconnecting
the endoscope.
2. Turn off the displaying unit and turn
it on again (Power off/Power on).
Still no image:
3. Refer to the Instructions for use
of the displaying unit for a detailed
troubleshooting guide or
alternatively take a new endoscope.
Replace the endoscope with a new one.
Return to live image by pressing the
blue live image tab or restart the
displaying unit by pressing the power
button for at least 2 seconds. When the
displaying unit is off, restart by
pressing power button once more.
15