5. Final Inspection and Test

CAUTIONS
To ensure the displaying unit functions optimally and to avoid compromising
electrical safety, always perform the tests specified in section 5 after it has been
reassembled. If any part of the tests fails after spare part replacement, do not use the
displaying unit.
Perform the following inspections and tests after assembly prior to use:
Examine displaying unit for visual damage
a)
Closely examine the displaying unit and other parts for any damage
displaying unit if it is damaged in any way.
b)
Visual inspection with a functioning scope
Connect the displaying unit to a suitable power outlet via the power cable
a functioning visualization device to the displaying unit by plugging it into the
visualization device connector (VDC) port
matching, and the arrows are aligned. Turn on the displaying unit on the power button.
Verify that a live video image appears on the screen by pointing the distal end of the
Ambu visualization device towards an object, e.g. the palm of your hand, when the Live
View
tab is selected in the toolbar to the left
I
III
For other relevant testing of the displaying unit, refer to the requirements of IEC 62353 or other
relevant safety tests according to local guidelines.
III
. Ensure that the connector colours are
IV
.
II
IV
I
. Do not use the
II
. Connect
15