AMC AMC-1225 Petunjuk Manual - Halaman 17

Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Sensor Keamanan AMC AMC-1225. AMC AMC-1225 17 halaman. Sensor modules

AMC-1225 SENSOR MODULES

6 MAINTENANCE

This section covers topics related to the maintenance of the AMC-1225 unit. A general
description of maintenance to be carried out is followed by a verification of operation and then
details about the sensor replacement.

6.1 GENERAL

The sensor unit should be brushed or wiped clean once a year or more of any dust or dirt which
settles on it depending on the accumulation.
The unit should NOT be submerged or placed under conditions where water or other liquids
would be able to enter the transmitter.

6.2 VERIFICATION OF OPERATION

Verification of operation should be performed at least once every 6 months. For highly
demanding applications more frequent verification is recommended.

6.3 SENSOR REPLACEMENT

Sensor life is typically in excess of 10 years. The sensor should be replaced when the green
LED turns OFF while the unit is powered.
CAUTION:
Turn off the power supply before attempting the following.
When the sensor needs replacing, reorder the Part Number listed in Product Information (page
4). To replace the sensor element remove the four screws and cover plate from the housing.
Unplug the used sensor element from its socket and discard. Then plug in the replacement
Figure 3-1
sensor element, refer to
for sensor polarity. Reinstall the cover plate and screws.
NOTE:
Allow 24 hours for the new sensor element to stabilize (burn-in) before recalibration, then
follow instructions in the Calibration section of this manual.
16