CO2 Meter RAD-0002-ZR Panduan Pengguna dan Panduan Instalasi - Halaman 12

Jelajahi secara online atau unduh pdf Panduan Pengguna dan Panduan Instalasi untuk Sistem Keamanan CO2 Meter RAD-0002-ZR. CO2 Meter RAD-0002-ZR 18 halaman. Oxygen o2 deficiency safety alarm
Juga untuk CO2 Meter RAD-0002-ZR: Manual (7 halaman)

CO2 Meter RAD-0002-ZR Panduan Pengguna dan Panduan Instalasi

Operation

Start-up

Verify that the RAD-0002-ZR is properly wired, and 12-24VDC power is being
supplied to the correct pin positions in the mating connector (see page 10). Check
all wiring connections before powering. After power-up, the monitor provides
accurate O2 measurements after a 10-minute warm up (Sensor requires brief
warm-up period).

Quick Test the System

1) Apply 100% Nitrogen to the senor port on the front of the monitor. Check that
all alarms trigger and the oxygen levels begin to drop.
2) AL1 will trigger Relay 1.
3) AL2 will trigger Relay 2.
4) AL3 will alarm.
5) The monitor will sound and flash (audible and visible) alarms.
6) If a CM-1027 Strobe is connected, it will flash.
***Allow 5 minutes for the monitor to come out of O2 alarm status.***

Unlocking/Locking Settings

1) Press and hold Up & Enter button for 10 seconds.
2) Press Mode until Aln is displayed.
3) Use Up/Down arrow to change lock to unlock.
4) Press Enter to save.
12 | P a g e
R A D - 0 0 0 2 - Z R / M a n u a l a n d I n s t a l l a t i o n
G u i d e