The Power (Green LED) will light continuously when the power is normally supplied.
If the device is powered by a battery, the Power (Green LED) will flash, and battery
indicator will appear and change with the battery voltage.
If the communication cable between the SEU & RDU is not connected securely, the
communication cable is loose or disconnected from the (RJ45) connectors, the fault
LED of SEU will blink as a reminder for the user to reconnect the cable. If the
communication cable is inserted into the wrong port on RDU, after about one
minute, the "Er7" will flash on the RDU LCD. Please plug the cables into the correct
ports on RDU and the unit will work normally.

RDU (Remote Display Unit)

The RDU displays the data from the SEU and provides visual and audible indication
that the SEU is in alarm status. The RDU is NOT an external/second sensor. The
RDU is connected to the SEU with a CAT5 cable. A 25-foot CAT5 cable is provided.
Users can source additional cable lengths as needed. The RDU should be placed
where it can be conveniently observed (eye level) before entering the room/space.
The "DIAG" function can test the communication between the SEU and RDU (see
9.1). Resetting the RAD-0102-6 CO2 Monitor is only available from SEU.

Strobes

The RAD-0102-6 can be equipped with strobes for additional visual alarms. Using
CAT5 cables, connect the strobes to the correct RJ45 ports on the SEU and RDU. If
CO2 levels exceed AL2, the strobe will flash. If the CO2 level exceeds AL3, the strobe
tempo increases. Please visit www.CO2Meter.com for strobe package. Part
Number: CM-1027.

Power

The RAD-0102-6 comes pre-wired with a 12V power supply that plugs in to a wall
outlet. The 12V power supply can be removed and 24V DC can be wired directly to
the device through the terminal block. Please use a 24V DC power converter to
properly convert your AC power source.
10 | P a g e
R A D - 0 1 0 2 - 6 / M a n u a l a n d I n s t a l l a t i o n