KEY WARNING MESSAGES
The CoaguChek automatically performs a system check to prevent
inaccurate results. If the meter detects a malfunction it displays the
"Error" message and appropriate symbol.
Error 3: Test Strip
Test strip is beyond it's expiry date. Check correct date and time is set on
meter. If correct, repeat test with new lot of test strips and code chip.
ccxs_user_EN.book Seite 74 Montag, 11. Juli 2005 7:11 19
Error 4: Test Strip Unusable
Often caused by applying the drop of blood before the countdown has
begun. Retest making sure you wait for the 180 second countdown before
Error messages
pricking the finger and applying the drop of blood.
Error messages overview
Error 5: Blood Application
Error message (always
prefaced by error)
Insufficient blood is applied to test strip. Repeat with new test strip.
Test strip
Error 6: Measurement
symbol
Knocking the meter or strip after applying the blood. Repeat with a new test
flashing
strip.
Temperature symbol
Error: Temperature symbol flashing
flashing
The meter is too cold (< 15˚C) or warm (> 32˚C) to measure correctly.
Battery symbol flashing
Cover symbol
flashing
CHANGE OF DATE AND TIME
IR symbol
flashing
When using the CoaguChek XS for the first time, or following a battery
QC symbol flashing
change, the date and time may need to be set.
Note: The meter will not work if the date and time are not set.
Code symbol
flashing
When the instrument is turned on it will automatically enter the set menu (if this
does not occur, press the set button on the left side of the monitor).
74
The set menu displays each setting option in the following sequence:
• Date Format select: 31/12/00 (DD/MM/YY)
• Date (year)
Press M until current year is displayed
• Date (month) Press M until current month is displayed
• Date (day)
Press M until current day is displayed
• Time format
select: 12 hour
• Time (hour)
Press M until correct hour is displayed
• Time (minute) Press M until correct minute is displayed Press set
Brief description
Refer
to
page
Test strip in monitor or
76
test strip unusable or the test
80
strip is not a CoaguChek XS PT
test strip
Monitor too warm or too cold
76
Battery level too low
78
Measurement chamber cover
78
open
Infrared download error
90
Quality control: test strip
86
QC
unusable
Code chip error
80
code
Press set
Press set
Press set
Press set
Press set
Press set
TOP TESTING TIPS
CoaguChek XS PT Test Strips
Some Points to Remember to
Ensure Accuracy of
ALWAYS wash your
Your INR Test Result
hands in warm water
and dry hands before
handling the test
strips.
Use the side of the
finger to obtain blood
- it's less painful.
After pricking the
finger gently milk it,
don't squeeze or press
as this may falsify the
coagulation values.
Apply blood to the clear
window on the test strip,
either side application or
top dosing.
Check the expiry date.
Don't use the strips after
the expiry date.
Always store strips in
the original container
and replace the lid
immediately.
Change your code key
each time you open a
new vial of strips.