Daavlin DT Home-Control Panduan Pengoperasian - Halaman 15

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Peralatan Medis Daavlin DT Home-Control. Daavlin DT Home-Control 19 halaman. Modes of operation - timed mode

Emissions Test
RF Emission Following
CISPR 11 (EN 55011)
Limits for harmonic current
emissions following IEC 61000-
3-2
Limitation of voltage changes,
voltage fluctuations and flicker
following IEC 61000-3-3
Emissions Test
Electrostatic discharge
immunity test following IEC
61000-4-2
Radiated, radio-frequency,
electromagnetic field
immunity test following IEC
61000-4-3
Electrical fast transient/burst
immunity test following IEC
61000-4-4
Surge immunity test following
IEC 61000-4-5
Conducted Immunity test
following IEC 61000-4-6
MNL-00049 [1]
Conformity
Class B
Class A
Compliant
Table 4 Electromagnetic Immunity
IEC 60601 Test Level
+/- 8kV (conductive
surfaces, coupling planes)
+/- 2kV, +/- 4kV, +/- 8kV,
and +/- 15kV (non-
conductive surfaces)
80 MHz to 2.7GHz @
10.0 V/m
+/- 2kV
+/- 0.5 kV, +/- 1.0 kV, +/-
2kV
0.15 MHz to 80 MHz @
3.0 Vrms
6.765 MHz to 6.795 MHz
@6.0 Vrms
13.553 MHz to 13.567
MHz @ 6.0 Vrms
26.057 MHz to 27.283
MHz @ 6.0 Vrms
40.66 MHz to 40.70 MHz
@ 6.0 Vrms
Proprietary and Confidential. Printed copies will not be updated.
EMC Environment Guide
The 1 Series device is suitable for
healthcare environment operation in
hospitals and clinics
The 1 Series device is suitable for
healthcare environment operation in
hospitals and clinics
The 1 Series device is suitable for
healthcare environment operation in
hospitals and clinic
Actual Level
+/- 8kV (conductive surfaces,
coupling planes)
+/- 2kV, +/- 4kV, +/- 8kV, and
+/- 15kV (non-conductive
surfaces)
80 MHz to 2.7GHz @ 10.0
V/m
+/- 2kV
+/- 0.5 kV, +/- 1.0 kV, +/- 2kV
0.15 MHz to 80 MHz @ 3.0
Vrms
6.765 MHz to 6.795 MHz
@6.0 Vrms
13.553 MHz to 13.567 MHz
@ 6.0 Vrms
26.057 MHz to 27.283 MHz
@ 6.0 Vrms
40.66 MHz to 40.70 MHz @
6.0 Vrms
Page 15 of 19